PPP-QLC-MFMR

804.NS.PPP-QLCMFMR
Silicon probe coated with low coercivity material for MFM in vacuum.

Each pack contains 10 probes. This product is not in stock but you can order and it will take around two weeks to get delivered
$ 1,280.00
Out of stock
Notify me when available


Description

f = 75 kHz   |   k = 2.8 N/m   |   tip coating: soft magnetic
Point Probe® Plus - Magnetic Force Microscopy - High Quality-Factor - Low Coercivity - Reflex Coating
The NANOSENSORS™ PPP-QLC-MFMR AFM probe combines the low disturbance of magnetic samples by a soft magnetic coating with the high mechanical quality factor under ultra high vacuum conditions of the Q30K-Plus-Series. Low coercivity and a Q-factor of more than 35,000 enable magnetic force microscopy of soft magnetic samples and high operation stability under UHV conditions. Due to the low coercivity of the tip coating the magnetisation of the tip will easily get reoriented by hard magnetic samples.

The soft magnetic coating on the tip has a coercivity of app. 0.75 Oe and a remanence magnetization of app. 225 emu/cm3 (these values were determined on a flat surface).
The SPM probe offers unique features:
  • soft magnetic coating on the tip side (coercivity of app. 0.75 Oe, remanence magnetization of app. 225 emu/cm3)
  • effective magnetic moment 0.75x of standard probes
  • guaranteed tip radius of curvature < 30 nm
  • magnetic resolution better than 35 nm
  • tip height 10 - 15 µm
  • Al coating on detector side of cantilever enhancing the reflectivity of the laser beam by a factor of about 2.5
  • excellent mechanical Q-factor under UHV conditions for high sensitivity
  • alignment grooves on backside of silicon holder chip
  • precise alignment of the cantilever position (within +/- 2 µm) when used with the Alignment Chip
  • compatible with PointProbe® Plus XY-Alignment Series

    As both coatings are almost stress-free the bending of the cantilever due to stress is less than 3.5% of the cantilever length. For enhanced signal strength the magnetization of the tip by means of a strong permanent magnet prior to the measurement is recommended.

Details

  • PPP-QLC-MFMR Specifications
  • Frequency (kHz):
    75.00
  • Max Frequency (kHz):
    115.00
  • Min Frequency (kHz):
    45.00
  • Spring Constant (N/m):
    2.80
  • Max Spring Constant (N/m):
    9.50
  • Min Spring Constant (N/m):
    0.52
  • Tip Radius (nm):
    23.00
  • Tip Height (um):
    12.50
  • Max Tip Height (um):
    15.00
  • Min Tip Height (um):
    10.00
  • Tip Shape:
    4 sided
  • Tip Material:
    Silicon
  • Tip Coating:
    Soft Magnetic
  • Reflective Coating:
    AI
  • Manufacturer:
    Nanosensors